Dexcom, Inc. 6340 Sequence Drive San Diego, CA 92121 USA 1.858.200.0200 1.877.339.2664 www.dexcom.comLBL012528 Rev 03 MT22690CONTINUOUS GLUCOSE

Pagina 2

8 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemGLOSSARY (continued from page before)Proles Sound pattern and volume level settings for

Pagina 3

98 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System1. Press the SELECT button to turn on the receiver. The 3-hour trend graph shows.2. Pr

Pagina 4

Dexcom G4 PLATINUM User’s Guide | 991. From the alerts menu, press the UP or DOWN button to select “High Alert” or “Low Alert” and press the SELECT

Pagina 5

100 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System95. To change this number, press the SELECT button, and then press the UP or DOWN butt

Pagina 6

Dexcom G4 PLATINUM User’s Guide | 1019.2.1 SETTING A SNOOZE TIME FOR YOUR HIGH AND LOW GLUCOSE ALERTS The snooze feature lets you delay your high an

Pagina 7

102 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemIf you set your fall rate to 2 mg/dL per minute and your sensor glucose readings fall a

Pagina 8 - GLOSSARY

Dexcom G4 PLATINUM User’s Guide | 1033. Press the UP or DOWN button to choose “On” or “Off.” Then, press the SELECT button to select “On” or “Off.”4

Pagina 9

104 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System9glucose readings. The out of range symbol in the upper right corner of the trend gra

Pagina 10

Dexcom G4 PLATINUM User’s Guide | 1054. Press the UP or DOWN button to choose the amount of time out of range after which the receiver will alert. P

Pagina 11

106 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System9Thesoft,normal,attentiveandHypoRepeatproleshavethefollowingalert sequence:

Pagina 12

Dexcom G4 PLATINUM User’s Guide | 107ALERT PROFILE OPTIONS (continued from page before)“HypoRepeat”prole:verysimilartothenormalprole,butit

Pagina 13 - SYSTEM DESCRIPTION

Dexcom G4 PLATINUM User’s Guide | 9GLOSSARY (continued from page before)System ReadingA sensor glucose reading shown on your receiver. This reading i

Pagina 14

108 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System9ALERT PROFILE DETAILS (continued from page before)PROFILE TYPEVIBRATE SOFT NORMAL ATTE

Pagina 15 - 1.1 SYSTEM CONTENTS:

Dexcom G4 PLATINUM User’s Guide | 1091. Press the SELECT button to turn on the receiver. The 3-hour trend graph will show.2. Press the SELECT butto

Pagina 16 - 1.2 INTRODUCTION

110 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemPAGE INTENTIONALLY LEFT BLANK

Pagina 17 - 1.3 SENSOR OVERVIEW

Dexcom G4 PLATINUM User’s Guide | 11110chapter tenEVENTS

Pagina 18 - 1.4 TRANSMITTER OVERVIEW

112 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System10

Pagina 19 - 1.5 RECEIVER OVERVIEW

Dexcom G4 PLATINUM User’s Guide | 113CHAPTER 10: EVENTSThe Events feature lets you record information that may help you and your healthcare professi

Pagina 20 - 1.6 SHARE OVERVIEW

114 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System4. Press the UP or DOWN button to choose the event you want: “Carbs,” “Insulin,” “Exer

Pagina 21 - SAFETY STATEMENT

Dexcom G4 PLATINUM User’s Guide | 1154. Press the SELECTbuttontoconrmthe entry.10.1.3 CARBOHYDRATESThe Carbs event lets you enter the amount o

Pagina 22

116 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System4. Press the LEFT or RIGHT button to choose either“OK”toconrmor“Cancel”todisca

Pagina 23 - 2.1 INDICATIONS FOR USE

Dexcom G4 PLATINUM User’s Guide | 1174. Press the LEFT or RIGHT button to choose either“OK”toconrmthisentryor“Cancel”to discard this entry,

Pagina 24 - 2.4 WARNINGS

10 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemPAGE INTENTIONALLY LEFT BLANK

Pagina 25

118 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System3. Press the UP or DOWN button to enter your exercise duration (0-360 minutes), and pr

Pagina 26 - 2.5 PRECAUTIONS

Dexcom G4 PLATINUM User’s Guide | 1192. Press the UP or DOWN button to choose your health event, and press the SELECT button.3. Check that the da

Pagina 27

120 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemPAGE INTENTIONALLY LEFT BLANK

Pagina 28

Dexcom G4 PLATINUM User’s Guide | 12111chapter elevenENDING A SENSOR SESSION

Pagina 29 - 2.6 CAUTION

122 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System11

Pagina 30

Dexcom G4 PLATINUM User’s Guide | 123CHAPTER 11: ENDING A SENSOR SESSIONYour sensor gives you sensor glucose readings for up to seven days. The perf

Pagina 31 - RISKS AND BENEFITS

124 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System11Youcansetthesealertswiththeprolessetting(seeChapter9,Section9.3.2,Aler

Pagina 32

Dexcom G4 PLATINUM User’s Guide | 125When you remove the sensor, make sure to pull out the sensor pod while the transmitter is still attached.1. Gen

Pagina 33 - 3.1 RISKS

126 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System1111.3 TRANSMITTER REMOVALDo not remove the transmitter while the sensor pod is still

Pagina 34 - 3.2 BENEFITS

Dexcom G4 PLATINUM User’s Guide | 12712chapter twelveTAKING CARE OF YOURDEXCOM G4 PLATINUM SYSTEM

Pagina 35 - 1, 2 , 3

Dexcom G4 PLATINUM User’s Guide | 111chapter oneDEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING (CGM) SYSTEM DESCRIPTION

Pagina 36

128 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System12

Pagina 37 - AND THE RECEIVER MAIN MENU

Dexcom G4 PLATINUM User’s Guide | 129CHAPTER 12: TAKING CARE OF YOUR DEXCOM G4 PLATINUM SYSTEM12.1 MAINTENANCETransmitter • Wipe the outside of t

Pagina 38

130 | Dexcom G4 PLATINUM Continuous Glucose Monitoring Systemformat is YYYY-MM-DD. Insert sensors on or before the end of the calendar day printed o

Pagina 39 - THE RECEIVER MAIN MENU

Dexcom G4 PLATINUM User’s Guide | 13113chapter thirteenTROUBLESHOOTING

Pagina 40 - FROM AN AC POWER OUTLET

132 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System13

Pagina 41

Dexcom G4 PLATINUM User’s Guide | 133CHAPTER 13: TROUBLESHOOTINGThischapterprovideshelpfultipsandinstructionstoxissuesyoumayhave while

Pagina 42 - 4.2 RECEIVER MENU OPTIONS

134 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System13 • I am not able to remove the transmitter latch: - Hold the sensor pod with one

Pagina 43

Dexcom G4 PLATINUM User’s Guide | 135 • See Chapter 7, Calibrating Your Dexcom G4 PLATINUM System, for further information.13.2.1 TYPES OF CALIBRA

Pagina 44

136 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System1313.2.2 WHAT TO DO FOR CALIBRATION PROMPTS1. When you see a calibration prompt, pres

Pagina 45

Dexcom G4 PLATINUM User’s Guide | 137SELECT button to clear the screen, this symbol shows in the status area.If you see this screen, wait at least

Pagina 46

12 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System1

Pagina 47 - WITH YOUR TRANSMITTER

138 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System13 • Make sure your sensor pod is not dislodged or peeling up. • Make sure your tran

Pagina 48

Dexcom G4 PLATINUM User’s Guide | 139If the difference between your sensor glucose reading and blood glucose value is greater than 20% of the blood g

Pagina 49 - AND DATE

140 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System13glucose to calibrate. Do not switch your meter in the middle of a sensor session. Blo

Pagina 50

Dexcom G4 PLATINUM User’s Guide | 141 • Make sure your transmitter is snapped in. • Make sure your sensor pod is not dislodged or peeling up. •

Pagina 51 - RECEIVER

142 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System132. Press the SELECT button with “OK” highlightedtoconrm.3. The Stop Sensor “thin

Pagina 52 - COMMUNICATION

Dexcom G4 PLATINUM User’s Guide | 143This screen means you recently turned Share “On,” and your receiver is having trouble connecting with your iPhon

Pagina 53 - CONTRAINDICATION

144 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System133. Charge your receiver.4. You may need to reset the time and date (see Chapter 4, S

Pagina 54

Dexcom G4 PLATINUM User’s Guide | 14513.9.3 TRANSMITTER LOW BATTERYThis screen shows when the transmitter nears the end of its battery life (see Cha

Pagina 55 - STARTING A SENSOR SESSION

146 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System13This screen means your receiver and transmitter are not communicating and you are not

Pagina 56

Dexcom G4 PLATINUM User’s Guide | 14713.11 ALERTS ARE NOT WORKING1. Make sure you have not turned off the sound and/or vibrations for the alerts. C

Pagina 57

Dexcom G4 PLATINUM User’s Guide | 13CHAPTER 1: DEXCOM G4® PLATINUM CONTINUOUS GLUCOSE MONITORING (CGM) SYSTEM DESCRIPTION1.1 SYSTEM CONTENTS: • sen

Pagina 58 - 6.1 BEFORE YOU START

148 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemPAGE INTENTIONALLY LEFT BLANK

Pagina 59 - PACKAGING

Dexcom G4 PLATINUM User’s Guide | 14914chapter fourteenTECHNICAL INFORMATION

Pagina 60

150 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14

Pagina 61 - 6.4 PLACING THE SENSOR

Dexcom G4 PLATINUM User’s Guide | 15116CHAPTER 14: TECHNICAL INFORMATION14.1 DEVICE PERFORMANCE CHARACTERISTICSNOTE: We recommend that you review t

Pagina 62 - 6.5 SENSOR INSERTION

152 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14and required insulin or oral medication to manage their diabetes. In the Original Stu

Pagina 63 - “ 2 CLICKS”

Dexcom G4 PLATINUM User’s Guide | 15316SW10586 SW10505 SW10050Includes Software 505 Incudes Software 505Does not include Software 505; includes Origi

Pagina 64 - 6.6 TRANSMITTER ATTACHMENT

154 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14Table 1. System Agreement to YSI within CGM Glucose RangesCGM Glucose Range1 (mg/dL)

Pagina 65

Dexcom G4 PLATINUM User’s Guide | 15516Agreement Relative to YSIAgreement between the System and blood glucose values is characterized using paired S

Pagina 66

156 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14(SW10050) or Software 505 (SW10505).Agreement When CGM Reads “LOW” or “HIGH”The Syste

Pagina 67 - 6.8 SENSOR STARTUP PERIOD

Dexcom G4 PLATINUM User’s Guide | 157System does not display glucose values below 40 mg/dL or above 400 mg/dL, the comparisons to the actual blood gl

Pagina 68 - 6.9 TAPING THE SENSOR POD

14 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemPRECAUTIONThe Dexcom G4 PLATINUM Sensor, Transmitter, and Receiver are not compatible wi

Pagina 69

158 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14Table 3-A. Concurrence of CGM Readings and YSI Values (Original Study)CGM (mg/dL)YSI

Pagina 70

Dexcom G4 PLATINUM User’s Guide | 159Table 3-B. Concurrence of CGM Readings and YSI Values (Software 505 Study)CGM (mg/dL)YSI (mg/dL)Row percentage

Pagina 71 - G4 PLATINUM SYSTEM

160 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14Another estimate used to show the accuracy of the System is the absolute percent diff

Pagina 72

Dexcom G4 PLATINUM User’s Guide | 16116Table 4. System Difference to YSI within CGM Glucose RangesCGMGlucose Ranges1 (mg/dL)Study2 Number of Paired

Pagina 73 - 7.1 CALIBRATION OVERVIEW

162 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14Low and High Glucose AlertsThe ability of the System to detect high and low glucose l

Pagina 74

Dexcom G4 PLATINUM User’s Guide | 16316Table 5. Hypoglycemic Alert EvaluationHypoglycemicAlert Level(mg/dL)Study1True Alert RateFalse Alert RateHypo

Pagina 75 - 7.2 HOW TO CALIBRATE

164 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14Table 6. Hyperglycemic Alert EvaluationHyperglycemicAlert Level(mg/dL)Study1True Ale

Pagina 76

Dexcom G4 PLATINUM User’s Guide | 165Hyperglycemia Alert RateThe Alert Rate shows how often the alert is right or wrong. The True Alert Rate is the %

Pagina 77

166 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14your alarm to sound 98% of the time and not to sound approximately 2% of time within

Pagina 78

Dexcom G4 PLATINUM User’s Guide | 167Table 7. Percentage of System Readings within YSI Values with Data Stratied in 2-Hour Increments after Calibra

Pagina 79 - 7.3 STARTUP CALIBRATION

Dexcom G4 PLATINUM User’s Guide | 151.3 SENSOR OVERVIEWThe sensor is the piece that comes in a sterile, sealed sensor pouch. The sensor is made up o

Pagina 80 - CALIBRATE

168 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14Table 8. Sensor Stability (Accuracy over Time)Day of Wear1Study2Number of paired CGM

Pagina 81

Dexcom G4 PLATINUM User’s Guide | 169Precision of System ReadingsIn the Original Study, 36 subjects wore two Systems. This was to look at how similar

Pagina 82

170 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemTable 9. Number of Readings Provided by Each Sensor Over 7-Days% of Total Possible Rea

Pagina 83 - AND TRENDS

Dexcom G4 PLATINUM User’s Guide | 17114Table 10. System Readings Within Wear DaysStatistic Study1Day 1 Day 2 Day 3 Day 4 Day 5 Day 6 Day 7All Days2M

Pagina 84

172 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14Table 11. System Agreement to SMBG Within CGM Glucose RangesCGM Glucose Ranges1(mg/d

Pagina 85

Dexcom G4 PLATINUM User’s Guide | 173Agreement and Accuracy Relative to SMBGDuring the study, agreement between the System and blood glucose values i

Pagina 86 - 8.1 SENSOR GLUCOSE READINGS

174 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14Table 12. CGM System Difference to SMBG within CGM Glucose RangesCGM Glucose Ranges1

Pagina 87

Dexcom G4 PLATINUM User’s Guide | 175Table 12 is categorized within CGM glucose ranges. Overall, the System in the Original Study reads, on average,

Pagina 88 - Which Trend Graph Do You See?

176 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System1414.2 PRODUCT SPECIFICATIONSSensor Product SpecicationsGlucose Range 40 - 400 mg/dLS

Pagina 89 - 8.2 RATE OF CHANGE ARROWS

Dexcom G4 PLATINUM User’s Guide | 177Moisture ProtectionIP28: temporary submersionProtection Against Electrical ShockType BF applied partTransmitter

Pagina 90 - No arrow

16 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System1.4 TRANSMITTER OVERVIEWThe transmitter is the gray, plastic “chip” that snaps into you

Pagina 91

178 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14device can withstand exposure to common electrostatic (ESD) and electromagnetic inter

Pagina 92

Dexcom G4 PLATINUM User’s Guide | 179Transmitter Electromagnetic Immunity Specications (continued from page before)Immunity Test IEC 60601 Test Leve

Pagina 93 - ALERTS, ALARM & PROFILES

180 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14Receiver Product SpecicationsPart Number MT22495Reading Frequency Every 5 minutesDim

Pagina 94

Dexcom G4 PLATINUM User’s Guide | 181Guidance and Manufacturer’s Declaration – Electromagnetic ImmunityThe receiver (P/N MT22495) is intended for use

Pagina 95 - 9.1 SETTING YOUR ALERTS

182 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14Receiver Electromagnetic Immunity Specications(continued from page before)Immunity T

Pagina 96

Dexcom G4 PLATINUM User’s Guide | 183System Electromagnetic Immunity SpecicationsImmunity Test IEC 60601 Test LevelReceiver Compliance LevelElectrom

Pagina 97

184 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14System Electromagnetic Immunity Specications (continued from page before)Immunity Te

Pagina 98 - 9.1.2.2 LOW GLUCOSE ALERT

Dexcom G4 PLATINUM User’s Guide | 185a.Fieldstrengthsfromxedtransmitters,suchasbasestationsforradio (cellular/cordless) telephones and l

Pagina 99 - 9.1.2.3 LOW GLUCOSE ALARM

186 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14Electromagnetic Emissions Specications (continued from page before)Emissions Test Co

Pagina 100

Dexcom G4 PLATINUM User’s Guide | 187Minimum Recommended Distance Between Transmitter and ReceiverRated MaximumOutput Power of Transmitter (W)Separa

Pagina 101

Dexcom G4 PLATINUM User’s Guide | 171.5 RECEIVER OVERVIEWThe receiver is the small hand-held device that looks like a cell phone. It is about 4 inc

Pagina 102 - 9.2 ADVANCED ALERTS

188 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14Power Supply/Charger SpecicationsDexcom P/N MT21255Class IIInput: AC Input 100-240

Pagina 103 - HIGH AND LOW GLUCOSE ALERTS

Dexcom G4 PLATINUM User’s Guide | 18914interference in a residential installation. This equipment generates uses and can radiate radio frequency ener

Pagina 104

190 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14PAGE INTENTIONALLY LEFT BLANK

Pagina 105

Dexcom G4 PLATINUM User’s Guide | 19115chapter fteenUSER ASSISTANCE

Pagina 106

192 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System15

Pagina 107 - 9.3 ALERT PROFILES

Dexcom G4 PLATINUM User’s Guide | 19317CHAPTER 15: USER ASSISTANCE Dexcom Website: www.dexcom.com Dexcom Address: 6340 Sequence Drive San Diego,

Pagina 108 - 9.3.1 ALERT PROFILE OPTIONS

194 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemPAGE INTENTIONALLY LEFT BLANK

Pagina 109 - 9.3.2 ALERT PROFILE DETAILS

Dexcom G4 PLATINUM User’s Guide | 19516chapter sixteenWARRANTY

Pagina 110

196 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System16

Pagina 111

Dexcom G4 PLATINUM User’s Guide | 197CHAPTER 16: WARRANTYDexcom G4 PLATINUM System Limited Warranty1. What is Covered and for How Long?Dexcom, Inc.

Pagina 113

18 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemYour receiver and transmitter wirelessly pair together to communicate securely and only

Pagina 114

198 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System16operation, maintenance, installation or adjustment; (f) installation, maintenance, an

Pagina 115 - 10.1 EVENTS

Dexcom G4 PLATINUM User’s Guide | 199AND DISCLAIMS ALL SUCH OTHER WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTY OF MERCHANTABILITY, FITNESS

Pagina 116 - AN EVENT

200 | Dexcom G4 PLATINUM Continuous Glucose Monitoring Systemthe original purchaser that the Dexcom G4 PLATINUM Transmitter is free from defects in

Pagina 117 - 10.1.3 CARBOHYDRATES

Dexcom G4 PLATINUM User’s Guide | 201of which was included with your Dexcom G4 PLATINUM CGM System and may be found at www.dexcom.com.3. What Are Dex

Pagina 118 - 10.1.4 INSULIN

202 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemPURPOSE, OR NON-INFRINGEMENT. EXCEPT TO THE EXTENT PROHIBITED BY APPLICABLE LAW, DEXCOM

Pagina 119 - 10.1.5 EXERCISE

Dexcom G4 PLATINUM User’s Guide | 20317chapter seventeenTRAVEL INFORMATION

Pagina 120 - 10.1.6 HEALTH

204 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System17

Pagina 121 - 10.2 DEXCOM STUDIO SOFTWARE

Dexcom G4 PLATINUM User’s Guide | 205CHAPTER 17: TRAVEL INFORMATIONIt is safe for you to go through the metal detector or be “handwanded” while wear

Pagina 122

206 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemPAGE INTENTIONALLY LEFT BLANK

Pagina 123 - ENDING A SENSOR SESSION

Dexcom G4 PLATINUM User’s Guide | 20718chapter eighteenAPPENDIX

Pagina 124

Dexcom G4 PLATINUM User’s Guide | 192chapter twoINDICATIONS FOR USE AND SAFETY STATEMENT

Pagina 125 - PRECAUTION

208 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System18

Pagina 126 - 11.2 REMOVING A SENSOR

Dexcom G4 PLATINUM User’s Guide | 209CHAPTER 18: APPENDIXAPPENDIX I, RECEIVER ALERTS, ALARM AND PROMPTSThe following tables describe the alarm, aler

Pagina 127

210 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System18Receiver Alerts, Alarm and Prompts (continued from page before)What will I see on the

Pagina 128 - 11.3 TRANSMITTER REMOVAL

Dexcom G4 PLATINUM User’s Guide | 211Receiver Alerts, Alarm and Prompts (continued from page before)What will I see on the receiver screen?Prompt, al

Pagina 129 - DEXCOM G4 PLATINUM SYSTEM

212 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System18Receiver Alerts, Alarm and Prompts (continued from page before)What will I see on the

Pagina 130

Dexcom G4 PLATINUM User’s Guide | 213Receiver Alerts, Alarm and Prompts (continued from page before)What will I see on the receiver screen?Prompt, al

Pagina 131 - 12.2 STORAGE

214 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System18Receiver Alerts, Alarm and Prompts (continued from page before)What will I see on the

Pagina 132 - 12.3 PRODUCT DISPOSAL

Dexcom G4 PLATINUM User’s Guide | 21520Receiver Alerts, Alarm and Prompts (continued from page before)What will I see on the receiver screen?Prompt,

Pagina 133 - TROUBLESHOOTING

216 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System18Receiver Alerts, Alarm and Prompts (continued from page before)What will I see on the

Pagina 134

Dexcom G4 PLATINUM User’s Guide | 217Receiver Alerts, Alarm and Prompts (continued from page before)What will I see on the receiver screen?Prompt, al

Pagina 135 - CHAPTER 13: TROUBLESHOOTING

20 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System2

Pagina 136

218 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemReceiver Alerts, Alarm and Prompts (continued from page before)What will I see on the r

Pagina 137

Dexcom G4 PLATINUM User’s Guide | 219Receiver Alerts, Alarm and Prompts (continued from page before)What will I see on the receiver screen?Prompt, al

Pagina 138 - 13.3 CALIBRATION ERROR

220 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System18Receiver Alerts, Alarm and Prompts (continued from page before)What will I see on the

Pagina 139 - 13.4 SYSTEM GLUCOSE ERROR

Dexcom G4 PLATINUM User’s Guide | 221Receiver Alerts, Alarm and Prompts (continued from page before)What will I see on the receiver screen?Prompt, al

Pagina 140 - 13.5 SENSOR INACCURACIES

222 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemAPPENDIX II, INDEXAAlarm, Low Glucose ...97Alerts ...

Pagina 141

Dexcom G4 PLATINUM User’s Guide | 223C Calibration ...71 Calibration Update ...78 How

Pagina 142

224 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemM Maintenance, System...129Menu Options...

Pagina 143

Dexcom G4 PLATINUM User’s Guide | 225 Insertion Site ... 58 Overview ...15 Pod ...

Pagina 144 - 13.7 SHARE PAIRING ERROR

226 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemLatch ... 9, 63 Overview ...

Pagina 145 - 13.8 BATTERY AND CHARGER

Dexcom G4 PLATINUM User’s Guide | 227APPENDIX III, SYMBOLS USED IN LABELINGThe following symbols may be found on the sensor, transmitter, and receive

Pagina 146 - 13.9.2 RECEIVER ERROR CODE

Dexcom G4 PLATINUM User’s Guide | 21CHAPTER 2: INDICATIONS FOR USE AND SAFETY STATEMENT2.1 INDICATIONS FOR USEThe Dexcom G4 PLATINUM Continuous Glu

Pagina 147 - 13.9.4 TRANSMITTER FAILED

228 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemSymbols (continued from page before)ManufacturerEC REPAuthorized Representative in the

Pagina 149 - 13.11 ALERTS ARE NOT WORKING

Dexcom, Inc. 6340 Sequence Drive San Diego, CA 92121 USA 1.858.200.0200 1.877.339.2664 www.dexcom.comLBL012528 Rev 03 MT22690CONTINUOUS GLUCOSE

Pagina 150

22 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System22.3 CONTRAINDICATIONS• Remove the Dexcom G4 PLATINUM Sensor, Transmitter, and Receive

Pagina 151 - TECHNICAL INFORMATION

Dexcom G4 PLATINUM User’s Guide | 23values may differ from sensor glucose readings. Using the sensor glucose readings for treatment decisions could l

Pagina 152

24 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System2• Sensor placement and insertion is not approved for sites other than the belly (abdom

Pagina 153 - CHARACTERISTICS

Dexcom G4 PLATINUM User’s Guide | 25to measure glucose. Insertion in those areas might affect sensor performance, and you might miss a low or high bl

Pagina 154

26 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System2ischangingatasignicantrate(e.g.,2-3mg/dL/minormore than 3 mg/dL each minute)

Pagina 155

Dexcom G4 PLATINUM User’s Guide | 27• The Dexcom G4 PLATINUM Sensor, Transmitter, and Receiver are not compatible with the SEVEN/SEVEN PLUS Transmit

Pagina 156

Dexcom G4 PLATINUM User’s Guide | 1USER’S GUIDEDEXCOM G4® PLATINUMCONTINUOUS GLUCOSE MONITORING SYSTEMRECEIVER WITH ShareTMIMPORTANT CONTACTS AND NU

Pagina 157 - Agreement Relative to YSI

28 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemPAGE INTENTIONALLY LEFT BLANK

Pagina 158

Dexcom G4 PLATINUM User’s Guide | 293chapter threeRISKS AND BENEFITS

Pagina 159

30 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System13

Pagina 160 - Values (Original Study)

Dexcom G4 PLATINUM User’s Guide | 31CHAPTER 3: RISKS AND BENEFITS3.1 RISKSThere are some known risks with using real-time CGM. You will not get sen

Pagina 161 - Accuracy Relative to YSI

32 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System3events happen, you might feel discomfort in the area the sensor is inserted.There is a

Pagina 162

Dexcom G4 PLATINUM User’s Guide | 33A1c values, reducing glycemic variability and time spent in low and high blood glucose ranges)1, 2 , 3 which can

Pagina 163

34 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemPAGE INTENTIONALLY LEFT BLANK

Pagina 164 - Low and High Glucose Alerts

Dexcom G4 PLATINUM User’s Guide | 354chapter fourCHARGING YOUR RECEIVER AND THE RECEIVER MAIN MENU

Pagina 165

36 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System4

Pagina 166 - The High Glucose Alert

Dexcom G4 PLATINUM User’s Guide | 37CHAPTER 4: CHARGING YOUR RECEIVER AND THE RECEIVER MAIN MENU4.1 CHARGING YOUR RECEIVER BATTERYThe receiver batt

Pagina 167

2 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemDEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEMtable of contentsGLOSSARY ...

Pagina 168 - Calibration Stability

38 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemCharge your receiver battery before each sensor insertion. Periodically check your batte

Pagina 169

Dexcom G4 PLATINUM User’s Guide | 395. The battery charging screen will show on the receiver.6. After a few seconds the trend graph will show with

Pagina 170 - Sensor Stability

40 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System44. After a few seconds, the trend graph screen will show the battery charging symbol i

Pagina 171 - Sensor Life

Dexcom G4 PLATINUM User’s Guide | 412Receiver Main Menu Options (continued from page before)Menu PurposeUser’s Guide ReferenceStart SensorTo start a

Pagina 172 - Number of Readings Provided

42 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemPAGE INTENTIONALLY LEFT BLANK

Pagina 173

Dexcom G4 PLATINUM User’s Guide | 435chapter veDEXCOM G4 PLATINUM SYSTEM SETUP

Pagina 174 - Glucose Ranges

44 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System5

Pagina 175

Dexcom G4 PLATINUM User’s Guide | 45CHAPTER 5: DEXCOM G4 PLATINUM SYSTEM SETUPThischapterhelpsyouwhenyourstsetupyourContinuousGlucoseMo

Pagina 176

46 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System5 (3) Press the SELECT button after you enter the last number or lettertoconrm

Pagina 177 - Adverse Events

Dexcom G4 PLATINUM User’s Guide | 475.2 THE SETTINGS MENUThe Settings menu lets you change the time, date or transmitter ID. The SetupWizardonlyw

Pagina 178 - 14.2 PRODUCT SPECIFICATIONS

Dexcom G4 PLATINUM User’s Guide | 3CHAPTER 5: DEXCOM G4 PLATINUM SYSTEM SETUP 435.1 Setting Up the Receiver and Pairing with Your Transmitter ...

Pagina 179

48 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System52. Press the RIGHT or SELECT button to highlight each value in the date and time.3.

Pagina 180 - Electromagnetic Immunity

Dexcom G4 PLATINUM User’s Guide | 491. From the Settings menu, press the UP or DOWN button to scroll to “Transmitter ID” and press the SELECT button

Pagina 181

50 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System55.3 CHECKING INFORMATION ABOUT YOUR DEXCOM G4 PLATINUM SYSTEMYou can check your receiv

Pagina 182

Dexcom G4 PLATINUM User’s Guide | 51PRECAUTIONThe transmission range from the transmitter to the receiver is up to 20 feet without obstruction. Wirel

Pagina 183

52 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System5Press the SELECT, LEFT or RIGHT button to see the trend graph. This antenna symbol show

Pagina 184

Dexcom G4 PLATINUM User’s Guide | 536chapter sixINSERTING A SENSOR AND STARTING A SENSOR SESSION

Pagina 185

54 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System6

Pagina 186

Dexcom G4 PLATINUM User’s Guide | 55CHAPTER 6: INSERTING A SENSOR AND STARTING A SENSOR SESSIONYou need a sensor, a transmitter, and a receiver to u

Pagina 187 - Electromagnetic Emissions

56 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System6.1 BEFORE YOU START• Make sure the correct transmitter ID has been entered into your

Pagina 188

Dexcom G4 PLATINUM User’s Guide | 576.2 REMOVING THE SENSOR FROM ITS PACKAGING• Wash your hands thoroughly, and dry them.• Carefully remove the sen

Pagina 189

4 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System8.3 Glucose Status Area Symbols ...89CHAPTER 9: ALERTS, ALARM

Pagina 190 - 14.3 FCC REQUIREMENTS

58 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System66.3 CHOOSING AN INSERTION SITEReview the tutorial disc for more help to learn how to i

Pagina 191

Dexcom G4 PLATINUM User’s Guide | 596.4 PLACING THE SENSOR1. Clean your skin at the sensor placement site with an alcohol wipe. Make sure the area

Pagina 192

60 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System63. Place the sensor horizontally, NOT vertically, on your skin. 4.Moveyourngersa

Pagina 193 - USER ASSISTANCE

Dexcom G4 PLATINUM User’s Guide | 612.Whilestillpinching,useyourotherhandtoplacetwongersabove the collar on the applicator barrel so th

Pagina 194

62 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System6 • Make sure the transmitter latch is down (against your body) before squeezing the t

Pagina 195 - SALES SUPPORT

Dexcom G4 PLATINUM User’s Guide | 63 c. With your other hand, pull the transmitter latch up and forward, over the transmitter, to snap it into plac

Pagina 196

64 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System63. Press the DOWN button to highlight “Start Sensor.” • The “Start Sensor” menu opti

Pagina 197 - WARRANTY

Dexcom G4 PLATINUM User’s Guide | 656.8 SENSOR STARTUP PERIODThe sensor needs a 2-hour startup period to adjust to being under the skin.Your trend g

Pagina 198

66 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System6time to calibrate your sensor. Chapter 7, Calibrating Your Dexcom G4 PLATINUM System, t

Pagina 199 - CHAPTER 16: WARRANTY

Dexcom G4 PLATINUM User’s Guide | 672.Conrmthatyouwanttoshutdownyour receiver. a. If you want to shut down, press the LEFT button to high

Pagina 200

Dexcom G4 PLATINUM User’s Guide | 512.2 Storage . ... 12912.3 Product Disposal . ...

Pagina 201

68 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemPAGE INTENTIONALLY LEFT BLANK

Pagina 202

Dexcom G4 PLATINUM User’s Guide | 697chapter sevenCALIBRATING YOUR DEXCOM G4 PLATINUM SYSTEM

Pagina 203

70 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System7

Pagina 204

Dexcom G4 PLATINUM User’s Guide | 71CHAPTER 7: CALIBRATING YOUR DEXCOM G4 PLATINUM SYSTEMYou must calibrate Dexcom G4 PLATINUM System sensor glucose

Pagina 205 - TRAVEL INFORMATION

72 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System7Example Minimum Calibration Schedule During Seven-Day Sensor SessionPRECAUTIONDo not us

Pagina 206

Dexcom G4 PLATINUM User’s Guide | 737.2 HOW TO CALIBRATEYou must enter the exact blood glucose value from your meter for each calibration. Blood glu

Pagina 207

74 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System7Only use blood glucose values between 40-400 mg/dL for calibration. If the blood glucos

Pagina 208

Dexcom G4 PLATINUM User’s Guide | 751. Wash and dry your hands, make sure your glucose test strips are not expired and have been stored properly, an

Pagina 209 - APPENDIX

76 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System77. Conrm that the blood glucose value you entered is correct. Entering incorrect valu

Pagina 210

Dexcom G4 PLATINUM User’s Guide | 777.3 STARTUP CALIBRATIONTwo hours after you start the sensor session (see Chapter 6, Section 6.7, Starting a Sens

Pagina 211 - CHAPTER 18: APPENDIX

6 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemGLOSSARYAlternative Site BG TestingThis is when you take a blood glucose value on your me

Pagina 212

78 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System7.4 12-HOUR CALIBRATION UPDATE Calibrateyoursystematleastevery12hoursafteryour

Pagina 213

Dexcom G4 PLATINUM User’s Guide | 79When you see this calibration prompt it means it is time to calibrate with a single blood glucose value.Take 1 bl

Pagina 214

80 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemPAGE INTENTIONALLY LEFT BLANK

Pagina 215

Dexcom G4 PLATINUM User’s Guide | 818chapter eightSENSOR GLUCOSE READINGS AND TRENDS

Pagina 216

82 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System1038

Pagina 217

Dexcom G4 PLATINUM User’s Guide | 83CHAPTER 8: SENSOR GLUCOSE READINGS AND TRENDS This chapter teaches you how to view your sensor glucose readings

Pagina 218

84 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System8WARNINGDo not use the Dexcom G4 PLATINUM System for treatment decisions, such as how mu

Pagina 219

Dexcom G4 PLATINUM User’s Guide | 85 • Each “dot” on the trend graph is a sensor glucose reading reported every 5 minutes. • The trend graph shows

Pagina 220

86 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System8Which Trend Graph Do You See?(Scroll up from the 3-hour graph to reach the 1-hour graph

Pagina 221

Dexcom G4 PLATINUM User’s Guide | 87The receiver displays “LOW” when the most recent sensor glucose reading is less than 40 mg/dL and “HIGH” when the

Pagina 222

Dexcom G4 PLATINUM User’s Guide | 7GLOSSARY (continued from page before)Dexcom Share™ SystemSecondary notication system to the Dexcom G4 PLATINUM Sy

Pagina 223

88 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System8Trend Arrows (continued from page before)Slowly rising: Your glucose is rising 1-2 mg/d

Pagina 224 - APPENDIX II, INDEX

Dexcom G4 PLATINUM User’s Guide | 89top of the trend graph, the trend arrows will not show. If the trend arrow is missing, but you are concerned that

Pagina 225

90 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System8Status Symbols (continued from page before)Glucose Reading ErrorThis symbol means the r

Pagina 226

Dexcom G4 PLATINUM User’s Guide | 919chapter nineALERTS, ALARM & PROFILES

Pagina 227

92 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System9

Pagina 228

Dexcom G4 PLATINUM User’s Guide | 93CHAPTER 9: ALERTS, ALARM & PROFILESThis chapter teaches you about your Dexcom G4 PLATINUM CGM System’s many

Pagina 229

94 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System9Default Alert/Alarm Settings (continued from page before)What will I see on the receive

Pagina 230

Dexcom G4 PLATINUM User’s Guide | 95Default Alert/Alarm Settings (continued from page before)What will I see on the receiver screen?What does this me

Pagina 231

96 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemWhen you have both your high and low alerts turned on, a gray zone on your trend graph s

Pagina 232

Dexcom G4 PLATINUM User’s Guide | 97the alert (to change your snooze settings see Chapter 9, Section 9.2, Advanced Alerts).9.1.2.3 LOW GLUCOSE ALARM